Complaint documentation
For this complaint code, please obtain the following:
- RGA (Return Goods Authorization)
Sensor complaint documentation
For this complaint code, please obtain the following:
- N/A
Problem
Receiver Screen displays "!HW" “ERR#” “ASSERT” Error Code or advises patient to "Call your local Dexcom representative." Receiver no longer displays glucose values & trends, and remains locked with "!HW" Error display.
After a soft rest, the issue might pass temporarily. Always replace display device with this issue if the device is in warranty.
Questions to be asked
N/A
Questions to be documented
114.1 What is the error code that was displayed on receiver?
114.2 What was the patient doing with the receiver when issue occurred?
Actions
- Choose one of the following:
• SW Assert
• HW_12C
• HW_RTC
• HW_RTC Timeloss
• HW PMIC Charger Timeout
• HW PMIC Charger Thermistor
• Other - Record the answer.
Images
n/aReasons behind
N/A
Related issues
N/A