YDC Product Troubleshooting Guide : 001 - Warmup takes longer than expected

Complaint documentation
  • Data
  • Receiver SN 
  • RGA (Return Goods Authorization)
Sensor complaint documentation
For this complaint code, please obtain the following:
  • Sensor insertion area
  • Sensor insertion date (if applicable)
  • Sensor lot number


Problem

After starting the session, the warm up period never ended, or the user was not prompted to enter required initial calibration, Readings have not been displayed.

G6/D1: If the user is in “No Cal” by using the sensor code, they will not receive the prompt.

If off-label insertion site is indicated, remind the patient of the approved site(s) and that if placed in other areas, the Dexcom System may not function properly.  Document this education in the Discussion/Resolution section.

Questions to be asked

N/A

Questions to be documented

001.1 How long did the warm up period last?

001.2 What was displayed during the warm up period?

Actions

  1. The warm up period may last beyond labeled duration. If the countdown indicator has not moved for the labeled duration, advise the user to stop the session and replace the sensor.
  2. If there was an alert, the warm up period may last longer than the labeled duration. Create an complaint for any alerts.  

Reasons behind

N/A

Related issues

N/A

Issues with same labels

There is no content with the specified labels


Back to overview

Comments:

Mercuriali, Massimo (MEMA7) bitte Sprachbox einfügen und Layout anpassen

Posted by SCNI6 at 03.05.2021 14:31