For this complaint code, please obtain the following:
- G6 Transmitter SN and Pump SN
- Landscape Mode (Screenshot
- RGA (Return Goods Authorization) : Report to Dexcom
Problem
Smart device displays: “Your transmitter was not found”
Questions to be asked
- Is the user using a compatible phone running a compatible operating system
- Is the user using the most up to date version of the app.
- Please ask the user to submit logs.
- Is the Transmitter attached to sensor and snapped firmly into place (snap it in firmly)
- Is the Transmiter connected to a live sensor (insert sensor if required)
- Have you acknowledged the pairing request on the phone
- What was displayed on the screen; was it ‘Now connecting to transmitter Wait 30min‘
- Is this the first time you are linking this Transmitter
- Is Bluetooth on the smart phone turned on
- Is Location on the smart phone turned on (device privacy or security settings)
- Are parental controls set on the smart phone; if YES confirm these do not prevent pairing
- Is the Transmitter linked with another app, e.g. Dexcom G6 app (unlink it or remove Dexcom G6 app from the smart phone)
Questions to be documented
- NOTE pump SN, Phone model & OS; Is phone model and OS on the compatibility list http://www.dexcom.com/compatibility; select country of residence and Dexcom G6 App
- NOTE Transmitter SN; was the Transmitter SN entered correctly
- NOTE Dexcom Username, if available
Actions
- Update the CamAPS FX App to the latest version available on Google Play
- On the smart phone, turn Bluetooth off and on
- Remove any transmitters no longer in use from the phone Bluetooth; unpair them by going to Settings ‑>Connections‑>Bluetooth
- On the smart phone, clear Bluetooth cache (Settings -> Apps->Show System Apps->Bluetooth->Storage->Clear cache)
- Power cycle the smart phone
- Re-pair the transmitter by going to the App Main Menu and selecting the transmitter Serial Number in blue; re-enter the Transmitter Serial Number
- If above fails, link the transmitter to Dexcom receiver (if available) or Dexcom G6 app (on the same and different phone) to check for transmitter fault; if unsuccessful the transmitter is likely to be at fault and should be replaced. Report to Dexcom (RGA)
- If spare transmitter available, try to pair it. If pairing successful, then the transmitter is at fault and needs to be replaced. Report to Dexcom (RGA)
- If spare transmitter not available, and transmitter had been working previously, replace the sensor and try pairing again
- If the second sensor fails, the transmitter is likely to be at fault and both sensor and transmitter should be replaced. Report to Dexcom (RGA)
Further options for bug fixes:
- General Connectivity Issues
- Huawei battery optimisation specific advice
- Android 10 specific advice
- Android 11 specific advice
TroubleShooting successful?: | Yes → Close as service ticket in C4C No → open Complaint to Dexcom |
---|---|
Complaint to open for: |
|
Addition: |
|
Complaint Coding: |
|
Error Code
Error Code | Internal Description | Message displayed |
DXC001 | MSG_UNKNOWN_ERROR | Internal error DXC001. Restart your smart device |
DXC002 | MSG_CYCLE_BLUETOOTH | Turn off and then turn on Bluetooth on your smart device |
DXC003 | MSG_CRC_CHECK | Internal error DXC003. Restart your smart device |
DXC004 | MSG_CRC_CALC | Internal error DXC004. Restart your smart device |
DXC005 | MSG_SN_NAME_WRONG | Incorrect transmitter SN entered. Re-enter the serial number … |
DXC007 | MSG_TRANS_TAKEN | Your transmitter may already be connected to another smart device |
DXC008 | MSG_TRANS_LOW_BATTERY | Your transmitter battery is critically low |
DXC009 | MSG_TRANS_FAILED | Your transmitter is not working. Pair with a new transmitter. |
DXC011 | MSG_TRANS_22DAYS | Your transmitter will expire in approximately 3 weeks |
DXC012 | MSG_TRANS_14DAYS | Your transmitter will expire in approximately 2 weeks |
DXC013 | MSG_TRANS_LAST_START | Your transmitter will last for one last sensor session |
DXC014 | MSG_TRANS_LAST_STOP | Your transmitter battery has died. Pair with a new transmitter |
DXC015 | MSG_SENSOR_6HOURS | Your sensor session will end in less than 6 hours |
DXC016 | MSG_SENSOR_2HOURS | Your sensor session will end in less than 2 hours |
DXC017 | MSG_SENSOR_30MIN | Your sensor session will end in less than 30 minutes |
DXC018 | MSG_SENSOR_EXPIRED | Your sensor session has ended |
DXC019 | MSG_URGENT_LOW | Urgent low glucose alarm |
DXC020 | MSG_LOW | Low glucose alert |
DXC021 | MSG_HIGH | High glucose alert |
DXC022 | MSG_FALL_RATE | Falling glucose alert |
DXC023 | MSG_RISE_RATE | Rising glucose alert |
DXC024 | MSG_SIGNAL_LOSS | Connection to your transmitter has been lost |
DXC025 | MSG_1ST_CALIB | Enter your 1st blood glucose reading |
DXC026 | MSG_2ND_CALIB | Enter your 2nd blood glucose reading |
DXC027 | MSG_NEXT_CALIB | Enter a new blood glucose reading to maintain your sensor |
DXC028 | MSG_NEXT_NEW_CALIB | Calibration error. Enter a new blood glucose reading. |
DXC029 | MSG_PLANNED_CALIB | Sensor recalibration required. In %min enter a new blood glucose reading |
DXC030 | MSG_SENSOR_FAILED | Sensor failure. You will not receive sensor glucose alerts |
DXC031 | MSG_PAIRING_FAILED | Your transmitter was not found |
DXC032 | MSG_PAIRING | Now connecting to transmitter |
DXC033 | MSG_PAIRED | Transmitter paired successfully |
DXC034 | MSG_SENSOR_WARM_UP | Sensor warming up. Wait for 2 hours |
DXC035 | MSG_START_SESS_ERROR | Internal error DX035. Restart your smart device |
DXC036 | MSG_END_SESS_ERROR | Internal error DX036. Restart your smart device |
DXC037 | MSG_CALIB_ERR | Internal error DX037. Restart your smart device |
DXC038 | MSG_INVALID_BG | Invalid blood glucose meter value |
DXC039 | MSG_CREATION_FAIL | Internal error DX039. Restart your smart device |
DXC040 | DB_DELETE_FAIL | Internal error DX040. Restart your smart device |
DXC041 | DB_INTEGRITY_FAIL | Internal error DX041. Restart your smart device |
DXC042 | DB_CONTEX_NULL | Internal error DX042. Restart your smart device |
DXC043 | DB_OPEN_EXCEPTION | Internal error DX043. Restart your smart device |
DXC044 | DB_GET_EXCEPTION1 | Internal error DX044. Restart your smart device |
DXC045 | DB_GET_EXCEPTION2 | Internal error DX045. Restart your smart device |
DXC046 | DB_INS_FAILED | Internal error DX046. Restart your smart device |
DXC047 | DB_INS_EXCEPTION | Internal error DX047. Restart your smart device |
DXC048 | DB_DEL_EXCEPTION | Internal error DX048. Restart your smart device |
DXC049 | DB_DEL_EXCEPTION2 | Internal error DX049. Restart your smart device |
DXC050 | DB_CIPHER | Internal error DX050. Restart your smart device |
DXC051 | END_OF_LIFE | Your transmitter battery has died. Pair with a new transmitter |
DXC052 | MSG_GATT_ERR_NOTI | Cannot establish reliable communication to the transmitter |
DXC053 | MSG_EXC_MSG_ACK | Internal error DX053. Restart your smart device |
DXC054 | MSG_RESTART_APP | Cannot establish reliable communication to the transmitter |
DXC055 | MSG_SENSOR_RESTART | Sensor can only be used for one session. Replace your sensor |
DXC056 | MSG_URGENT_LOW_SOON | Urgent low soon alert |
DXC057 | DB_DEL_EXCEPTION3 | Internal error DX057. Restart your smart device |
DXC058 | MSG_NO_READINGS | No sensor glucose readings alert. You will not receive any alerts … |
DXC059 | MSG_SCAN_STUCK | Internal error DX059. Restart your smart device |
Important
As soon as a troubleshooting was not successful you have to open a complaint (Dexcom Portal) and you have to tell the patient to submit the logfile, a screenshot in Landscape Mode, the time of the error message (time / date), the pump SN, the G6 transmitter SN and Dexcom username.
Images
Reasons behind
- Only one app, Dexcom G6 app or CamAPS app, can connect to the Dexcom transmitter at any one time, while Dexcom G6 receiver and CamAPS app can be connected at the same time
Related issues
Issues with same labels
There is no content with the specified labels
Attachments: