For this complaint code, please obtain the following:
- Pump SN
- Logfile
- Landscape Mode (Screenshot)
Problem
Smart device displays:
“Signal loss”, “No sensor reading”,
“Sensor error. Check in 10 minutes”, “Sensor failed”
Questions to be asked
- Is the user using a compatible phone running a compatible operating system
- Is the user using the most up to date version of the app.
- Please ask the user to submit logs.
- Are you using a recommended smart device and OS? https://www.diabetescare.abbott/support/manuals.html
- Is the Bluetooth on your smart device turned on? (if YES, turn Bluetooth Off and On again; go to phone Settings > Connections > Bluetooth)
- Was the phone within 6 meters (19 feet) when loss of data occurred; move the phone within range
- Was the phone away from a microwave, induction hob, WiFi booster, Bluetooth headset, baby monitor or other “electronic noise”; move the phone away from these devices
- Were you using any other Bluetooth accessories at the time of signal loss
- Was the message “No sensor reading, check in 10min” displayed? (if YES, this is likely to be a temporary sensor issue, please wait 10 minutes)
- Was the message “Sensor failed” displayed? (if YES, replace the sensor; Refer to Abbott Diabetes Care for further information
Questions to be documented
- NOTE pump SN, Phone model & OS;
- What length of time was the data loss; NOTE duration
Actions
- Make sure there are no obstructions (walls, water, body mass) or interference (microwave, WiFi booster, baby monitor, induction hob) or other Bluetooth accessories
- If signal loss is still displayed perform troubleshooting:
- On the smart phone, turn Bluetooth off and on
- On the smart phone, clear Bluetooth cache (Settings -> Apps -> Show System Apps -> Bluetooth -> Storage -> Clear cache) (only possible up to Android OS 12, for OS 13 -> directly restart smartphone)
- On the smart phone, turn off battery optimization (Settings -> Apps -> Show System Apps -> Bluetooth -> Battery -> Unrestricted)
- On the smart phone, clear CamAPS FX app cache (Settings -> Apps -> CamAPS FX -> Storage -> Clear cache)
- Check if mylife CamAPS FX battery optimization is “Unrestricted” (Settings -> Apps -> CamAPS FX -> Battery
-> Unrestricted) - Check if connection to FreeStyle Libre 3 sensor can be established
- If no, Power cycle the smart phone
- Check if connection to FreeStyle Libre 3 sensor can be established
- Restart the smart device and wait 10 minutes
- Restart the sensor: go to Menu > Start new sensor, rescan the sensor with the back of the smart device
- If still no data, perform troubleshooting documented in “unable to start FreeStyle Libre 3“ section
Trouble Shooting successful? | Yes → Close as service ticket in C4C No → Was customer already transferred from Abbott ADC to YPS?
|
---|---|
Complaint to open for: |
|
Addition: | Complaint Management reports case to 2nd level support |
Complaint Coding: | F118: CamAPS App A527: CGM: Signal loss – permanent A526: CGM: Signal loss – frequent |
Error Codes
LBR001 | MSG_UNKNOWN_ERROR | Internal error LBR001. Restart your smart device |
LBR002 | MSG_CRC_CHECK | Internal error LBR002. Restart your smart device |
LBR003 | MSG_CRC_CALC | Internal error LBR003. Restart your smart device |
LBR004 | TOO_BIG_GAP | Internal error LBR004. Restart your smart device |
LBR005 | MSG_WAKELOCK_ERROR | Internal error LBR005. Restart your smart device |
LBR006 | MSG_BAD_IDX_1 | Internal error LBR006. Restart your smart device |
LBR007 | MSG_BAD_IDX_2 | Internal error LBR007. Restart your smart device |
LBR008 | MSG_BAD_IDX_3 | Internal error LBR008. Restart your smart device |
LBR009 | MSG_BAD_IDX_4 | Internal error LBR009. Restart your smart device |
LBR010 | MSG_EXC_SAVE_INFO | Internal error LBR010. Restart your smart device |
LBR011 | MSG_START_SESS_ERR | Internal error LBR011. Restart your smart device |
LBR012 | DB_CREATION_FAIL | Internal error LBR012. Restart your smart device |
LBR013 | DB_DELETE_FAIL | Internal error LBR013. Restart your smart device |
LBR014 | DB_INTEGRITY_FAIL | Internal error LBR014. Restart your smart device |
LBR015 | DB_CONTEX_NULL | Internal error LBR015. Restart your smart device |
LBR016 | DB_OPEN_EXCEPTION | Internal error LBR016. Restart your smart device |
LBR017 | DB_GET_EXCEPTION1 | Internal error LBR017. Restart your smart device |
LBR018 | DB_GET_EXCEPTION2 | Internal error LBR018. Restart your smart device |
LBR019 | DB_INSULIN_FAILED | Internal error LBR019. Restart your smart device |
LBR020 | DB_INSULIN_EXCEPTION | Internal error LBR020. Restart your smart device |
LBR021 | DB_DEL_EXCEPTION | Internal error LBR021. Restart your smart device |
LBR022 | DB_DEL_EXCEPTION2 | Internal error LBR022. Restart your smart device |
LBR023 | HANDLER_STUCK | Internal error LBR023. Restart your smart device |
LBR024 | MSG_EXC_MSG_ACK | Internal error LBR024. Restart your smart device |
LBR025 | DB_DEL_EXCEPTION3 | Internal error LBR025. Restart your smart device |
LBR050 | MSG_SENSOR_6HOURS | Your sensor will end in less than 6 hours |
LBR051 | MSG_SENSOR_2HOURS | Your sensor will end in less than 2 hours |
LBR052 | MSG_SENSOR_30MINUTES | Your sensor will end in less than 30 minutes |
LBR053 | MSG_SENSOR_EXPIRED | Your sensor ended. You will not receive sensor glucose alerts |
LBR054 | MSG_URGENT_LOW | Urgent low glucose alarm |
LBR055 | MSG_LOW | Low glucose alert |
LBR056 | MSG_HIGH | High glucose alert |
LBR057 | MSG_FALL_RATE | Falling glucose alert |
LBR058 | MSG_RISE_RATE | Rising glucose alert |
LBR059 | MSG_SIGNAL_LOSS | Connection to your sensor has been lost. You will not receive sensor glucose alerts |
LBR060 | MSG_SENSOR_FAILED | Replace sensor. Your sensor is not working |
LBR061 | MSG_SENSOR_WARM_UP | Sensor warming up. Wait for 60 minutes |
LBR062 | MSG_URGENT_LOW_SOON | Urgent low glucose soon alert |
LBR063 | MSG_NO_READINGS | No sensor glucose readings alert. You will not receive sensor glucose alerts |
LBR064 | MSG_TEMP_LOW | Sensor too cold |
LBR065 | MSG_TEMP_HIGH | Sensor too hot |
LBR066 | MSG_CURR_SENSOR | The sensor you just scanned is already started |
LBR067 | MSG_WRONG_PIN | Unable to connect to sensor. Go to main menu, stop the previous sensor |
LBR068 | MSG_VITAMIN_C | Do not take high doses of vitamin C (more than 500mg per day) |
Important
As soon as a troubleshooting was not successful, you have to open a complaint and you have to tell the patient to submit the logfile, a screenshot in Landscape Mode, the time of the error message (time / date) and the pump SN.
Images
Reasons behind
- Use of incompatible smart device / phone / operating system (OS)
- Phone too far away or high level of electronic noise
- Sensor stopped communicating with the smart device due to sensor error, message “Sensor failed”
- Temporary sensor error or sensor failure with message “No sensor reading, check in 10 minutes”
- The sensor insertion technique or positioning of the sensor could be the cause. The instructions in the app Help should be followed.
FreeStyle Libre 3 Error List
Please refer the customer to Abbott Diabetes Care for the error codes listed below.
Code | Description | Possible Cause(s) | Interface | Display | App |
---|---|---|---|---|---|
365 | Replace Sensor Error | Sensor in Error State This error is logged once, immediately when Patch Failure is received via BLE from the Patch. If this corresponds to Patch event 11 (Algo error termination), then the corresponding extended error code is displayed, and Patch event 4011 will additionally be logged. If it corresponds to another error termination, then no extended error code will be displayed, and the corresponding Patch event 4xxx will additionally be logged. This may also be logged during activation when scanning a Sensor in error state | NFC BLE | Replace sensor screen Replace sensor system alarm | C |
366 | Sensor Fallout Error | The sensor has come off the patient – reinsert if possible This error is logged once, immediately when Insertion Detection Failure is received via BLE from the Patch. This corresponds to Patch Event 5, and will be logged along with 4005. | BLE | Sensor fallout screen | |
367 | Incompatible Sensor Type Error | The sensor scanned during Activation is not compatible with this Receiver – try the correct sensor type | NFC | Incompatible sensor screen | |
372 | Sensor Not Yours Error | Another sensor already in use by another Receiver has been scanned during Activation. | NFC | Sensor already in use screen | |
373 | Current Glucose Result DQ Error | The Current Glucose reading that was Viewed was determined by the algorithm to be invalid. Note: This is logged only when user Views a DQ reading, at most once per Sensor Life Count. | BLE | Results screen with Sensor Error | C |
380 | ESA is detected ESA marks the first 12 hours of data as having Invalid Data DQ errors. | Sensor reading is unusually low. Logged when the first Current glucose reading is received with ESA via BLE. Only logged once per Sensor. Note: If user is out of range of the Sensor during the 12 hours of ESA, this will not be logged. | BLE | Results screen with Sensor Error | |
4001 – 4xxx | Patch Reported Errors/Events | Patch Error Codes 1 – xxx This corresponds to error/event codes received from the Patch. The Patch error code (1-xxx) is prepended with a 4, logged in the database, and displayed to the user. Note: Not all Patch Errors will be displayed. Errors 1-4 and 6-7 are not stored or displayed in the event log. Patch Error 11 (4011) will be accompanied by 365, and Patch Error 5 (4005) will be accompanied by 366. See Patch Error spec for more information. |