YDC Product Troubleshooting Guide : 003SL - APP NOT RECEIVING FREESTYLE LIBRE 3 DATA FOR LONGER THAN 20 MINUTES

Complaint documentation

For this complaint code, please obtain the following:

  • Pump SN
  • Logfile
  • Landscape Mode (Screenshot)


Problem

Smart device displays:

“Signal loss”, “No sensor reading”,
“Sensor error. Check in 10 minutes”, “Sensor failed”

Questions to be asked

  • Is the user using a compatible phone running a compatible operating system
  • Is the user using the most up to date version of the app.
  • Please ask the user to submit logs.
  1. Are you using a recommended smart device and OS? https://www.diabetescare.abbott/support/manuals.html
  2. Is the Bluetooth on your smart device turned on? (if YES, turn Bluetooth Off and On again; go to phone Settings > Connections > Bluetooth)
  3. Was the phone within 6 meters (19 feet) when loss of data occurred; move the phone within range
  4. Was the phone away from a microwave, induction hob, WiFi booster, Bluetooth headset, baby monitor or other “electronic noise”; move the phone away from these devices
  5. Were you using any other Bluetooth accessories at the time of signal loss
  6. Was the message “No sensor reading, check in 10min” displayed? (if YES, this is likely to be a temporary sensor issue, please wait 10 minutes)
  7. Was the message “Sensor failed” displayed? (if YES, replace the sensor; Refer to Abbott Diabetes Care for further information

Questions to be documented

  1. NOTE pump SN, Phone model & OS;
  2. What length of time was the data loss; NOTE duration

Actions

  • Make sure there are no obstructions (walls, water, body mass) or interference (microwave, WiFi booster, baby monitor, induction hob) or other Bluetooth accessories
  • If signal loss is still displayed perform troubleshooting:
  • On the smart phone, turn Bluetooth off and on
  • On the smart phone, clear Bluetooth cache (Settings -> Apps -> Show System Apps -> Bluetooth -> Storage -> Clear cache) (only possible up to Android OS 12, for OS 13 -> directly restart smartphone)
  • On the smart phone, turn off battery optimization (Settings -> Apps -> Show System Apps -> Bluetooth -> Battery -> Unrestricted)
  • On the smart phone, clear CamAPS FX app cache (Settings -> Apps -> CamAPS FX -> Storage -> Clear cache)
  • Check if mylife CamAPS FX battery optimization is “Unrestricted” (Settings -> Apps -> CamAPS FX -> Battery
    -> Unrestricted)
  • Check if connection to FreeStyle Libre 3 sensor can be established
  • If no, Power cycle the smart phone
  • Check if connection to FreeStyle Libre 3 sensor can be established
  • Restart the smart device and wait 10 minutes
  • Restart the sensor: go to Menu > Start new sensor, rescan the sensor with the back of the smart device
  • If still no data, perform troubleshooting documented in “unable to start FreeStyle Libre 3“ section


Trouble Shooting successful?

Yes → Close as service ticket in C4C

No Was customer already transferred from Abbott ADC to YPS?

  • If No→ Close as service ticket in C4C and transfer customer to AbbottDiabetes Care
  • If Yes → Create Complaint

Complaint to open for:

  • Ypsomed 
  • Product: mylife CamAPS FX
  • Material number: 700021738

Addition:

Complaint Management reports case to 2nd level support

Complaint Coding:

F118: CamAPS App

A527: CGM: Signal loss – permanent
(Error can only be solved with de- and re-installation, re-pairing)

A526: CGM: Signal loss – frequent
(Error can only be solved with workarounds (e.g. emptied cache))

 

Error Codes

LBR001

MSG_UNKNOWN_ERROR

Internal error LBR001. Restart your smart device

LBR002

MSG_CRC_CHECK

Internal error LBR002. Restart your smart device

LBR003

MSG_CRC_CALC

Internal error LBR003. Restart your smart device

LBR004

TOO_BIG_GAP

Internal error LBR004. Restart your smart device

LBR005

MSG_WAKELOCK_ERROR

Internal error LBR005. Restart your smart device

LBR006

MSG_BAD_IDX_1

Internal error LBR006. Restart your smart device

LBR007

MSG_BAD_IDX_2

Internal error LBR007. Restart your smart device

LBR008

MSG_BAD_IDX_3

Internal error LBR008. Restart your smart device

LBR009

MSG_BAD_IDX_4

Internal error LBR009. Restart your smart device

LBR010

MSG_EXC_SAVE_INFO

Internal error LBR010. Restart your smart device

LBR011

MSG_START_SESS_ERR

Internal error LBR011. Restart your smart device

LBR012

DB_CREATION_FAIL

Internal error LBR012. Restart your smart device

LBR013

DB_DELETE_FAIL

Internal error LBR013. Restart your smart device

LBR014

DB_INTEGRITY_FAIL

Internal error LBR014. Restart your smart device

LBR015

DB_CONTEX_NULL

Internal error LBR015. Restart your smart device

LBR016

DB_OPEN_EXCEPTION

Internal error LBR016. Restart your smart device

LBR017

DB_GET_EXCEPTION1

Internal error LBR017. Restart your smart device

LBR018

DB_GET_EXCEPTION2

Internal error LBR018. Restart your smart device

LBR019

DB_INSULIN_FAILED

Internal error LBR019. Restart your smart device

LBR020

DB_INSULIN_EXCEPTION

Internal error LBR020. Restart your smart device

LBR021

DB_DEL_EXCEPTION

Internal error LBR021. Restart your smart device

LBR022

DB_DEL_EXCEPTION2

Internal error LBR022. Restart your smart device

LBR023

HANDLER_STUCK

Internal error LBR023. Restart your smart device

LBR024

MSG_EXC_MSG_ACK

Internal error LBR024. Restart your smart device

LBR025

DB_DEL_EXCEPTION3

Internal error LBR025. Restart your smart device

LBR050

MSG_SENSOR_6HOURS

Your sensor will end in less than 6 hours

LBR051

MSG_SENSOR_2HOURS

Your sensor will end in less than 2 hours

LBR052

MSG_SENSOR_30MINUTES

Your sensor will end in less than 30 minutes

LBR053

MSG_SENSOR_EXPIRED

Your sensor ended. You will not receive sensor glucose alerts

LBR054

MSG_URGENT_LOW

Urgent low glucose alarm

LBR055

MSG_LOW

Low glucose alert

LBR056

MSG_HIGH

High glucose alert

LBR057

MSG_FALL_RATE

Falling glucose alert

LBR058

MSG_RISE_RATE

Rising glucose alert

LBR059

MSG_SIGNAL_LOSS

Connection to your sensor has been lost. You will not receive sensor glucose alerts

LBR060

MSG_SENSOR_FAILED

Replace sensor. Your sensor is not working

LBR061

MSG_SENSOR_WARM_UP

Sensor warming up. Wait for 60 minutes

LBR062

MSG_URGENT_LOW_SOON

Urgent low glucose soon alert

LBR063

MSG_NO_READINGS

No sensor glucose readings alert. You  will not receive sensor glucose alerts

LBR064

MSG_TEMP_LOW

Sensor too cold

LBR065

MSG_TEMP_HIGH

Sensor too hot

LBR066

MSG_CURR_SENSOR

The sensor you just scanned is already started

LBR067

MSG_WRONG_PIN

Unable to connect to sensor. Go to main menu, stop the previous sensor

LBR068

MSG_VITAMIN_C

Do not take high doses of vitamin C (more than 500mg per day)

Important

As soon as a troubleshooting was not successful, you have to open a complaint and you have to tell the patient to submit the logfile, a screenshot in Landscape Mode, the time of the error message (time / date) and the pump SN.

Images

Reasons behind

  • Use of incompatible smart device / phone / operating system (OS)
  • Phone too far away or high level of electronic noise
  • Sensor stopped communicating with the smart device due to sensor error, message “Sensor failed”
  • Temporary sensor error or sensor failure with message “No sensor reading, check in 10 minutes”
  • The sensor insertion technique or positioning of the sensor could be the cause. The instructions in the app Help should be followed.

FreeStyle Libre 3 Error List

Please refer the customer to Abbott Diabetes Care for the error codes listed below.

CodeDescriptionPossible Cause(s)InterfaceDisplayApp
365Replace Sensor ErrorSensor in Error State This error is logged once, immediately when Patch Failure is received via BLE from the Patch. If this corresponds to Patch event 11 (Algo error termination), then the corresponding extended error code is displayed, and Patch event 4011 will additionally be logged. If it corresponds to another error termination, then no extended error code will be displayed, and the corresponding Patch event 4xxx will additionally be logged. This may also be logged during activation when scanning a Sensor in error state

NFC

BLE

Replace sensor screen

Replace sensor system alarm

C

366Sensor Fallout ErrorThe sensor has come off the patient – reinsert if possible This error is logged once, immediately when Insertion Detection Failure is received via BLE from the Patch. This corresponds to Patch Event 5, and will be logged along with 4005.BLESensor fallout screen
367Incompatible Sensor Type ErrorThe sensor scanned during Activation is not compatible with this Receiver – try the correct sensor typeNFCIncompatible sensor screen
372Sensor Not Yours ErrorAnother sensor already in use by another Receiver has been scanned during Activation.NFCSensor already in use screen
373Current Glucose Result DQ ErrorThe Current Glucose reading that was Viewed was determined by the algorithm to be invalid. Note: This is logged only when user Views a DQ reading, at most once per Sensor Life Count.BLEResults screen with Sensor ErrorC
380ESA is detected ESA marks the first 12 hours of data as having Invalid Data DQ errors.Sensor reading is unusually low. Logged when the first Current glucose reading is received with ESA via BLE. Only logged once per Sensor. Note: If user is out of range of the Sensor during the 12 hours of ESA, this will not be logged.BLEResults screen with Sensor Error
4001 – 4xxxPatch Reported Errors/Events

Patch Error Codes 1 – xxx

This corresponds to error/event codes received from the Patch. The Patch error code (1-xxx) is prepended with a 4, logged in the database, and displayed to the user. Note: Not all Patch Errors will be displayed. Errors 1-4 and 6-7 are not stored or displayed in the event log. Patch Error 11 (4011) will be accompanied by 365, and Patch Error 5 (4005) will be accompanied by 366. See Patch Error spec for more information.